Heal Force: HFsafe cabinets, HF CO2 incubators, Neofuge centrifuges and water purification systems
Technical selection of Heal Force laboratory equipment (Shanghai, China): HFsafe and Alphasafe biological safety cabinets, AlphaClean and OptiClean clean benches, HF CO2 incubators, Neofuge centrifuges and water purification systems.
HFsafe biological safety cabinets
HFsafe 900/1200/1500/1800 with a manual or motorized sash, and HFsafe-LC 900LC to 1800LC. Both recirculate 70% of the air and exhaust 30%, with 0.35 m/s downflow and 0.53 m/s inflow. HFsafe has a 10° tilt; HFsafe-LC adds a scheduled UV lamp start.
- Recirculation/exhaust: 70/30%
- Downflow: 0.35 m/s
- Inflow: 0.53 m/s
- HFsafe tilt: 10°
HFsafe-TE and HFsafe-LC total exhaust cabinets
Versions exhausting 100% of the air: HFsafe-TE and the total exhaust HFsafe-LC. Downflow is 0.30 m/s and inflow 0.53 m/s, with a 10° tilt and upper and lower sash alarms. The HFsafe-TE exhaust duct can be linked to building ventilation so supply and exhaust adjust automatically.
- Exhaust: 100%
- Downflow: 0.30 m/s
- Inflow: 0.53 m/s
- Tilt: 10°
AlphaClean 1300 and OptiClean 1300 clean benches
Vertical laminar flow benches for two operators on one side. AlphaClean 1300, launched in 2016, has a 304 stainless worktop, prefilter, airflow sensor, 5° tilt and UV lamp timer. OptiClean 1300 has a one-piece filter with life monitoring, a counterweighted sash and a worktop from 1400 mm.
- Airflow: vertical
- AlphaClean tilt: 5°
- OptiClean worktop: from 1400 mm
- OptiClean width: up to 1500 mm
HF90, HF151, HF212, HF240 and HF160W CO2 incubators
HF90, HF151, HF212 and HF240 are air-jacketed, with 151, 151, 212 and 240 L chambers, ambient +5 °C to 50 °C and 0–20% CO2. HF90 and HF240 add 90 °C moist heat decontamination. HF160W is water-jacketed, with a 184 L chamber, up to 55 °C and an infrared CO2 sensor.
- Chambers: 151, 212 and 240 L
- CO2: 0–20%
- HF90, HF240: 90 °C decontamination
- HF160W: 184 L chamber
HF100 multi-gas and HF180 CO2 incubators
HF100 is a 151 L multi-gas incubator: ambient +3 °C to 55 °C, a zirconium oxide O2 sensor with an optional 1–85% range, automatic gas cylinder switchover. HF180 is a CO2 incubator with 180 °C dry heat sterilisation in under 2 hours, an 8-inch touchscreen, USB and 4–20 mA outputs.
- HF100: 151 L chamber
- HF100: 1–85% O2 (option)
- HF180: 180 °C sterilisation
- HF180: 4–20 mA output
Neofuge 13, 13R, 15 and 15R centrifuges
Benchtop high-speed centrifuges; R models are refrigerated. Neofuge 13: up to 13,800 rpm, 24×1.5/2.0 mL rotor. Neofuge 13R: up to 15,000 rpm, 24×1.5/2.0 mL rotor, optional 16×5 mL, −10 to +40 °C. Neofuge 15 and 15R: up to 16,000 rpm, loads up to 4×100 mL, the 15R up to 20,800×g.
- Neofuge 13: up to 13,800 rpm
- Neofuge 13R: up to 15,000 rpm
- Neofuge 15, 15R: up to 16,000 rpm
- Neofuge 15R: up to 20,800×g
Neofuge 1600, 1600R and 23R centrifuges
Benchtop centrifuges for larger loads. Neofuge 1600 and 1600R run up to 15,000 rpm and take up to 4×400 mL, a choice of 8 rotors and 10 stored programs. Neofuge 23R is refrigerated from −20 to +40 °C, takes 0.2 mL up to 4×180 mL, has a Strong Drive spindle and stores 9 programs.
- Neofuge 1600, 1600R: up to 15,000 rpm
- Neofuge 1600, 1600R: up to 4×400 mL
- Neofuge 23R: up to 4×180 mL
- Neofuge 23R: −20 to +40 °C
Smart, Easy and PureForce ROE water purification systems
Easy delivers three water grades at once: pure, high-purity and ultrapure. Smart Plus has a 30 L tank and a standard leak sensor. PureForce ROE is rated at 70 and 100 L/h; floor-standing PureForce ROE units have 60 and 350 L tanks.
- Easy: three water grades
- Smart Plus tank: 30 L
- PureForce ROE: 70 and 100 L/h
- PureForce ROE tank: 60 or 350 L
Choosing a cabinet or clean bench: HFsafe, HFsafe-TE, Alphasafe, AlphaClean
HFsafe and HFsafe-LC cabinets recirculate 70% of the air and exhaust 30%, with 0.35 m/s downflow and 0.53 m/s inflow. HFsafe comes as 900, 1200, 1500 and 1800 models with a manual or motorized sash.
HFsafe-TE and the total exhaust HFsafe-LC discharge 100% of the air at 0.30 m/s downflow; the HFsafe-TE duct can be linked to the building ventilation.
Alphasafe has front glass at least 6 mm thick and a 20 cm working opening, and drops to low speed when the sash is closed.
AlphaClean 1300 and OptiClean 1300 clean benches provide vertical flow for two operators working from one side; the OptiClean 1300 worktop is at least 1400 mm wide within an outer width of up to 1500 mm.
HF CO2 incubators: chamber volume, jacket, decontamination
HF90, HF151, HF212 and HF240 are air-jacketed: 151, 151, 212 and 240 L, ambient +5 °C to 50 °C, 0–20% CO2. HF90 and HF240 have 90 °C moist heat decontamination, and HF240 can take 6 small inner doors. HF160W is water-jacketed: 184 L, up to 55 °C, infrared CO2 sensor.
HF100 is a 151 L multi-gas incubator with a zirconium oxide O2 sensor and an optional 1–85% range, and automatic gas cylinder switchover. HF180 sterilises with 180 °C dry heat in under 2 hours and outputs data over USB and a 4–20 mA analogue output.
Neofuge centrifuges and water purification systems
For microtubes there are Neofuge 13 with a 24×1.5/2.0 mL rotor (up to 13,800 rpm) and the refrigerated Neofuge 13R with a 24×1.5/2.0 mL rotor and an optional 16×5 mL rotor (up to 15,000 rpm, −10 to +40 °C). Neofuge 15 and 15R run up to 16,000 rpm; their maximum load is 4×100 mL.
For larger volumes, Neofuge 1600 and 1600R take up to 4×400 mL with a choice of 8 rotors, and Neofuge 23R takes up to 4×180 mL refrigerated from −20 to +40 °C.
The Easy system delivers pure, high-purity and ultrapure water at the same time, and Smart Plus works with a 30 L tank. PureForce ROE is rated at 70 and 100 L/h; for central water supply in a laboratory there are floor-standing PureForce ROE units with 60 and 350 L tanks.
Heal Force and model names are trademarks of their respective owners and are used only to identify the equipment for which we source parts. SpareX is not an official dealer, distributor or representative of Heal Force and is not affiliated with the trademark owner. Origin, documents and warranty are stated for each item in the quotation.
Related equipment manufacturers
- Equipment and parts for Infors HT
- Equipment and parts for Biotage
- Equipment and parts for Biolin Scientific
Documents and sources
- Our story — group structure
- After sales support — laboratory equipment service
- HFsafe manual sash cabinet — overview and company address
- HFsafe motorized sash cabinet — overview
- HFsafe-LC cabinets — series overview
- HFsafe cabinets — airflow data
- HFsafe-LC cabinets — airflow data
- HFsafe-TE cabinets — total exhaust
- HFsafe-LC total exhaust cabinets — airflow data
- Alphasafe cabinets — overview
- Clean benches — model list
- AlphaClean 1300 clean bench — features
- OptiClean 1300 clean bench — features
- HF90 incubator — specifications
- HF240 incubator — specifications
- HF151 incubator — specifications
- HF212 incubator — specifications
- HF160W incubator — specifications
- HF100 multi-gas incubator — overview
- HF100 incubator — specifications
- HF180 incubator — specifications
- Neofuge 13 centrifuge — specifications
- Neofuge 13R centrifuge — overview
- Neofuge 13R centrifuge — specifications
- Neofuge 15R centrifuge — overview
- Neofuge 15 centrifuge — specifications
- Neofuge 15R centrifuge — specifications
- Neofuge 1600 centrifuge — specifications
- Neofuge 1600R centrifuge — specifications
- Neofuge 23R centrifuge — overview
- Neofuge 23R centrifuge — specifications
- Water purification — model list
- Smart water system — modules
- Smart Plus water system — modules
- Easy water system — modules
- PureForce ROE compact floor unit — overview
- PureForce ROE mid-size floor unit — overview
- PureForce ROE — output