MDC Precision Vacuum Valves, Feedthroughs and Viewports

MDC Precision (USA): gate valves, CF and ISO flanges, ceramic-to-metal power and coaxial feedthroughs, breaks, viewports, motion feedthroughs, bell jars and chambers.

C-Loc and Uni-Loc Gate Valves

MDC circular gate valves come with .625″ to 12″ ports, manual or electropneumatic actuators and CF, ISO-KF, ISO-LF or ANSI ASA flanges. The body is 300 series stainless, the bellows AM-350, and the gate locks with a C-Loc or Uni-Loc mechanism. Each valve is helium leak tested.

  • Ports: .625–12″
  • Leak test: 2×10⁻¹⁰ cc/sec He
  • Air supply: 70–100 psig
  • Circular valve solenoid: 120 VAC, 4 W

Power and High Voltage Feedthroughs

MDC ceramic-to-metal feedthroughs are made by its Ceramic Seals unit. The single conductor high voltage feedthrough with a stainless steel conductor is rated up to 30,000 V DC and 1 A; the UHV version bakes to 450 °C, the HV version to 150 °C. The water-cooled FHC-1000 carries up to 1000 A at 50 V.

  • High voltage: to 30,000 V DC
  • High voltage current: to 1 A
  • High voltage UHV: to 450 °C
  • FHC-1000: to 1000 A

Coaxial and Multipin Feedthroughs

BNC coaxial feedthroughs are rated 500 V DC, MHV and SHV up to 5000 V DC. BNC and MHV look alike but must never be cross-mated; on SHV the male pin sits in the feedthrough. The multipin instrumentation feedthrough is rated up to 700 V DC and 10 A with a MIL-C-5015 mating connector.

  • BNC: 500 V DC
  • MHV and SHV: to 5000 V DC
  • Multipin: to 700 V DC, 10 A
  • Multipin connector: MIL-C-5015

Electrical Breaks and RF Feedthroughs

MDC ceramic-to-metal break adapters are rated 10 or 15 kV DC and envelopes 8 or 15 kV DC, with 1×10⁻¹⁰ atm cc/sec helium leak tightness. The UHV version has one rotatable and one nonrotatable CF flange and Kovar sleeves. The single tube RF power feedthrough handles up to 35 kW at up to 13.5 MHz.

  • Breaks: 10 and 15 kV DC
  • Envelopes: 8 and 15 kV DC
  • RF, single tube: to 35 kW
  • RF frequency: to 13.5 MHz

Glass, Quartz and Sapphire Viewports

MDC glass viewports fit CF, ISO-KF and ISO-LF flanges with clear apertures from 5/8 to 5-3/8 inch; CF glass ones bake to 400 °C; gradients above 10 °C per minute must be avoided. Quartz viewports are not heated above 200 °C. Sapphire CF viewports bake repeatedly to 450 °C, ISO ones to 200 °C.

  • Glass CF: to 400 °C
  • Quartz: to 200 °C
  • Sapphire CF: to 450 °C
  • Glass apertures: 5/8–5-3/8″

CF, 316LN and ISO-LF Flanges

MDC 304 stainless CF flanges are UHV rated: the 2-1/8″ DN25CF and 2-3/4″ DN40CF flanges bake to 450 °C, and the 2-1/8″ size comes rotatable or nonrotatable. 316LN CF flanges offer ultra-low magnetic permeability. For high vacuum lines there are ISO-LF flanges such as NW400.

  • 2-3/4″ DN40CF: to 450 °C
  • 2-1/8″ DN25CF: to 450 °C
  • 316LN CF flanges: UHV
  • ISO-LF flange: NW400

Rotary Feedthroughs and Manipulators

MDC standard rotary motion feedthroughs provide precise rotation in vacuum, while differentially pumped rotary feedthroughs deliver high torque in UHV. The basic Wobble Stick feedthrough gives 4.5″ (114.3 mm) of linear travel and 22° of angular motion.

  • Wobble Stick travel: 4.5″ (114.3 mm)
  • Wobble Stick angle: 22°
  • Differential pumping: UHV
  • Standard: rotary motion

Bell Jars, Baseplates and Chambers

MDC stainless steel bell jars seal with an FKM gasket in a dovetail groove and carry a 4″ Pyrex viewport. Feedthrough collars offer 4, 8 or 18 ports with 2-3/4″ CF or ISO-KF NW40 flanges; baseplates suit 12, 18 and 24″ chambers. Custom chambers are built from a sketch or DWG and DXF drawings.

  • Collars: 4, 8, 18 ports
  • Baseplates: 12, 18, 24″
  • Bell jar viewport: Pyrex 4″
  • Drawings: DWG, DXF

Selecting MDC Precision Vacuum Components

The MDC Precision range covers flanges and fittings, gate and other valves, viewports, electrical and fluid feedthroughs, motion feedthroughs, bell jars and custom chambers.

Electrical feedthroughs, breaks, sapphire and quartz viewports are made by the Ceramic Seals business unit in Sarasota, Florida; MDC and Insulator Seal (ISI) joined forces in 1998.

Voltage, current and temperature are chosen per series: BNC coaxial feedthroughs are rated 500 V DC, MHV and SHV 5000 V DC, and BNC must never be cross-mated with MHV. MDC catalogue dimensions are given in inches, so the flange is checked against the part number.

Gate Valve and Viewport Maintenance

On MDC gate valves the carriage and gate mechanism come out through the bonnet flange for seal replacement and cleaning without removing the valve body from the system. A gasket kit holds one bonnet and one gate seal.

UHV valves with a copper bonnet seal and a Kalrez 4079 gate seal bake to 250 °C open and 200 °C closed.

Glass viewports with scratches or nicks should be replaced, and thermal gradients above 10 °C per minute must be avoided. Quartz viewports are not heated above 200 °C.

Industrial equipment

Request a quote

Describe your requirements and attach up to 5 files, each up to 20 MB.

You can also drag photos or documents here.

You can also email us. Opening your email app does not send an enquiry automatically.

WeChat · Nastia

English-speaking contact

WeChat QR code: Nastia

Scan the QR code to add Nastia on WeChat.

Save QR image

On this phone: save the image, then choose it from your photos in WeChat’s scanner.

Telegram · Nastia

Contact Nastia

Telegram QR code: @ananasprosto

Scan the QR code or open @ananasprosto in Telegram.@ananasprosto в Telegram.

Save QR image

Scan the QR code or open @ananasprosto in Telegram.