Novasina water activity meters and sensors
Technical selection of Novasina instruments: LabMaster-aw neo, LabTouch-aw, LabSwift-aw and LabStart-aw water activity meters, nSens-HT humidity sensors, nSens-DP and nSens-P pressure sensors, nLink+ transmitters.
LabMaster-aw neo water activity meter
The LabMaster-aw neo measures 0.030 to 1.000 aw with 0.0001 aw resolution and full chamber thermostatting at 0 to 60 °C. Accuracy is ±0.0030 aw in the adjustment range, and fast mode gives a result within 10 minutes. Salt standards are read by RFID; the audit trail meets 21 CFR 11.
- Range 0.030 to 1.000 aw
- Chamber 0 to 60 °C
- Accuracy ±0.0030 aw
- RFID salt standards
LabTouch-aw water activity meter
The LabTouch-aw measures 0.030 to 1.000 aw with ±0.005 aw accuracy; the chamber is heated only. Fast mode gives a result within 10 minutes. The CM-2 cell serves regular samples, the CM-3 samples with more than 0.1 vol% ethanol, and acid, volatile and Redox filters are available.
- Accuracy ±0.005 aw
- Chamber heating only
- CM-2 or CM-3 cell
- Weight 2.8 kg
LabSwift-aw portable water activity meter
The LabSwift-aw takes one measurement in the lab and the next on the line, with an optional rechargeable battery. It covers 0.030 to 1.000 aw with ±0.010 aw accuracy and has no temperature control. The maker lists powders and dried fruits as typical samples.
- Accuracy ±0.010 aw
- Optional battery
- No temperature control
- Weight 1.2 kg
LabStart-aw entry-level water activity meter
The LabStart-aw is built for samples in the low to medium aw range. It uses the Novasina resistive electrolytic sensor, covers 0.03 to 1.00 aw with 0.01 aw resolution and has no temperature control. Adjustment uses two points, 33 and 75 %, and one SAL-T 75 % standard is supplied.
- Range 0.03 to 1.00 aw
- Resolution 0.01 aw
- Adjustment at 33 and 75 %
- SAL-T 75 % standard
nSens-HT-EIS, EIH and CSP humidity sensors
Digital humidity and temperature sensors: the nSens-HT-EIS with an electrolytic element and ±0.4 % typical accuracy, the heated nSens-HT-EIH for condensing environments with ±0.8 %, and the capacitive nSens-HT-CSP with ±1.5 %. Calibration data is stored in the sensor itself.
- nSens-HT-EIS: ±0.4 % rH
- nSens-HT-EIH: heated, ±0.8 %
- nSens-HT-CSP: ±1.5 % rH
- Temperature −20 to +80 °C
nSens-DP and nSens-P pressure sensors
The nSens-DP measures air differential pressure: nSens-DP-AE for ±25 Pa, nSens-DP-BE for ±250 Pa and nSens-DP-CE for ±5000 Pa. The nSens-P-ME measures absolute pressure from 280 to 1260 hPa with ±1 hPa accuracy at 20 °C. The sensors plug in and need no separate power supply.
- nSens-DP-AE: ±25 Pa
- nSens-DP-BE: ±250 Pa
- nSens-DP-CE: ±5000 Pa
- nSens-P-ME: 280 to 1260 hPa
nLink+ DI, IP and EC transmitters
The nLink+ comes as the DI with a 2.8 inch touch display, the IP in an IP65 housing and the EC for rail mounting in a cabinet. Outputs are 0/4 to 20 mA or 0/2 to 10 V, Modbus RTU or two relays; some models have a built-in dP sensor. The IP AS and DI AS take 5 to 48 V DC or 24 V AC.
- 4 to 20 mA or 0 to 10 V
- Modbus RTU output
- IP AS: 5 to 48 V DC
- Sensor cable up to 100 m
Measuring cells and protective filters
The LabTouch-aw uses CM-2 and CM-3 cells, the LabMaster-aw neo uses awSens ENS and awSens ELS; the CM-3 and ELS resist ethanol. Acid, volatile and Redox filters protect the electrolytic element from chemical contamination. The regeneration kit is only for the CM-2 and awSens ENS.
- CM-3 cell: P/N 2600536
- awSens ELS: P/N 2601615
- Regeneration kit 2601962
- Acid, volatile, Redox filters
Choosing a Novasina water activity meter
Novasina AG of Lachen, Switzerland, has developed instruments for water activity, humidity and air pressure since 1958. All LAB series instruments are built on the proprietary resistive-electrolytic sensor.
The LabMaster-aw neo fully thermostats the chamber at 0 to 60 °C, reaches ±0.003 aw and keeps a 21 CFR 11 audit trail; data goes to LIMS as XML, it prints to an Epson TM-U220D and weighs about 10 kg.
The heated LabTouch-aw gives ±0.005 aw, the portable LabSwift-aw without temperature control gives ±0.01 aw, and the LabStart-aw serves samples in the low to medium aw range.
Humidity and differential pressure in cleanrooms
For cleanrooms, storage and HVAC, nSens-HT and nSens-DP sensors work with nLink+ transmitters. Outputs of 4 to 20 mA, 0 to 10 V, Modbus RTU or relays connect to PLC and BMS systems, and the differential pressure transmitter comes with one or two sensors.
The capacitive nSens-HT-CSP has an ideal range of 20 to 80 % rH, and the heated nSens-HT-EIH serves condensing environments. The maker notes that in pharma storage at 70 % rH and above mold risk rises, and that differential pressure control supports ISO 14644 and EU GMP Annex 1.
Cells, filters and standards
The measuring cell follows the sample: CM-2 or awSens ENS for regular samples, CM-3 or awSens ELS when ethanol exceeds 0.1 vol%. Acid, volatile and Redox filters protect the element but lengthen the measurement.
Instruments are calibrated with reusable SAL-T salt standards: the LabMaster-aw neo has 11 adjustment points and 7 standards supplied, the LabTouch-aw 8 points and 4 standards, the LabSwift-aw 7 points and 3 standards, the LabStart-aw 2 points and 1 standard.
Novasina and model names are trademarks of their respective owners and are used only to identify the equipment for which we source parts. SpareX is not an official dealer, distributor or representative of Novasina and is not affiliated with the trademark owner. Origin, documents and warranty are stated for each item in the quotation.
Related equipment manufacturers
- Equipment and parts for HIRANUMA
- Equipment and parts for KOFLOC
- Equipment and parts for TOHO Electronics
- Equipment and parts for Shinko Technos
Documents and sources
- Novasina AG — site notice
- Water activity — instrument overview
- LabMaster-aw neo — specifications
- LabTouch-aw — specifications
- LabSwift-aw — specifications
- LabStart-aw — specifications
- Humidity monitoring — sensor overview
- nSens-HT-EIS — specifications
- nSens-HT-EIH — specifications
- nSens-HT-CSP — specifications
- Differential pressure — overview
- nSens-DP — specifications
- nSens-P — specifications
- nLink+ family — documents
- nLink+ IP AS — data sheet
- nLink+ DI AS — data sheet
- Transmitter for automation — outputs
- Relay and analog transmitters — overview
- Filter selection — data sheet