Sensonics vibration sensors and protection systems

Technical selection of Sensonics equipment: Senturion XPR04 and XPR08 proximity probes, DNX8031 transmitters, PZS accelerometers, PZDC and VEL-G velocity sensors, Sentry G3 and DN26 G3 monitors, CS25 air gap sensors.

Senturion XPR04 and XPR02 eddy current probes

A Senturion X system of probe, extension cable and driver measures shaft vibration and position. The 8 mm tip XPR04 covers 2.5 mm at 7.87 mV/µm or 4.0 mm at 3.94 mV/µm, and the 5 mm tip XPR02 covers up to 2.5 mm. The probe operates to +180 °C and the driver to +90 °C.

  • Tip diameter 8 mm or 5 mm
  • Sensitivity 7.87 mV/µm
  • Probe to +180 °C, driver to +90 °C
  • System length 5, 7 or 9 m

XPR08, XPR12 and XPD disc probes for turbines

The 20 mm tip XPR08 measures up to 8.0 mm at 1.97 mV/µm for shaft position and rotor expansion on steam turbines. The XPR12 and the XPD12 disc probe with a 25 mm tip measure up to 12 mm at 1.38 mV/µm. The 50 mm tip XPD25 disc probe measures up to 25 mm at 0.63 mV/µm.

  • XPR08: up to 8.0 mm, 1.97 mV/µm
  • XPR12 and XPD12: up to 12 mm
  • XPD25: up to 25 mm, 0.63 mV/µm
  • Disc mount for limited space

DNX8031, DNX8033 and DNX8038 4–20 mA transmitters

The DNX8031 combines the proximity probe driver with signal processing and outputs 4–20 mA for peak to peak shaft vibration up to 0–250 µm. The DNX8033 outputs shaft position or thrust from −1.0 to +1.0 mm, and the DNX8038 outputs speed from 1 Hz to 10 kHz. Signals go directly to the PLC or DCS.

  • Shaft vibration up to 0–250 µm
  • Thrust −1.0 to +1.0 mm
  • Speed 1 Hz to 10 kHz
  • 4–20 mA current output

PZS shear mode accelerometers

The PZS accelerometer mounts on the bearing housing and outputs an IEPE signal from 100 to 500 mV/g, with a ±70 g peak range. Response is 0.4 Hz to 10 kHz, it operates from −40 to +140 °C, and the welded stainless steel body has an IP68 sealed connector. An Ex ia IIC T4 Ga version is available.

  • Sensitivity 100–500 mV/g
  • Response 0.4 Hz – 10 kHz
  • Operates −40 to +140 °C
  • Marking Ex ia IIC T4 Ga

PZV, PZDC and VEL-G velocity sensors

The PZV outputs an IEPE signal of 4 mV/mm/s over 2 Hz to 6 kHz. The PZDC outputs 4–20 mA for velocity up to 50 mm/s over 10 Hz to 1 kHz. The VEL-G moving coil sensor is self generating at 4 to 20 mV/mm/s over 4.5 Hz to 2 kHz, and the loop powered VEL-GDC operates to +100 °C.

  • PZV: 4 mV/mm/s, IEPE
  • PZDC: 4–20 mA up to 50 mm/s
  • VEL-G: self generating coil
  • VEL-GDC: operates to +100 °C

Sentry G3 API 670 machinery protection monitors

Each Sentry G3 module has 4 independently programmed channels for vibration, thrust, rotor expansion, temperature, speed, eccentricity and air gap. A module carries 4 alarm relays (expandable to 12) and 8 analogue outputs; a 19″ 3U rack holds up to 24 channels with dual redundant power.

  • 4 channels per module, API 670
  • 4 relays (to 12), 8 outputs
  • Up to 24 channels in 3U
  • Redundant power and Modbus

DN26 G3 protection modules and DN2601–DN2612 monitors

The DIN rail DN26 G3 module monitors 2 channels of bearing vibration, shaft vibration or thrust position, and a third channel measures speed or acts as a phase reference. Three alarm relays expand to seven. Dedicated DN2601, DN2602 and DN2603 monitors cover bearing, shaft and thrust.

  • 2 channels plus speed channel
  • 3 relays, expandable to 7
  • Modbus RS485 and TCP/IP
  • Pumps, fans, motors

CS capacitive air gap sensors for hydro generators

CS sensors measure the rotor to stator air gap of hydro generators: CS07 covers 7 mm, CS25 covers 25 mm at 400 mV/mm and 640 µA/mm, and CS75 covers 75 mm. The flat CS25 sensor mounts on the stator wall, operates from −20 to +125 °C and tolerates magnetic fields up to 1.5 T.

  • Ranges from 7 to 75 mm
  • CS25: 400 mV/mm, 640 µA/mm
  • Operates −20 to +125 °C
  • System length 5 or 10 m

Choosing Sensonics sensors for each measuring point

For shaft vibration and position on fluid film bearings Sensonics makes Senturion X eddy current systems: XPR04 and XPR02 for shaft vibration, XPR08, XPR12 and XPD disc probes for thrust and rotor expansion.

Where the signal goes straight to a PLC, DNX8031, DNX8033 and DNX8038 transmitters give a 4–20 mA output.

Bearing housing vibration is measured by PZS accelerometers and PZV, PZDC and VEL-G velocity sensors. For valve and component position there are LVDT transducers from 2.5 to 600 mm, and the DIN rail DN8032 converts their signal to 4–20 mA.

Turbine protection systems: Sentry G3, DN26 G3 and overspeed

Sentry G3 covers the full set of API 670 turbine supervisory measurements in one rack, while DN26 G3 DIN rail modules serve smaller machines such as pumps, fans, motors and centrifuges.

The Sensonics overspeed protection system uses 2 out of 3 voting, responds in less than 20 ms and is built to IEC 61508 SIL-3. For seismic protection the triaxial SA-3 seismic switch has an actuating level from 0.005 to 0.5 g.

Proximity probe installation and channel wiring

The probe gap is set by the voltage at the 3.5 mm driver socket, and the 5, 7 or 9 m system length is selected by a switch. Replacing the probe, cable or driver keeps the interchangeability error within ±5%; the ECPH holder lets the probe be removed and refitted with no further adjustment.

For PZDC loop powered sensors the cable screen is earthed at the monitor end only; the recommended loop is 24 V with a 250 ohm load.

Industrial equipment

Request a quote

Describe your requirements and attach up to 5 files, each up to 20 MB.

You can also drag photos or documents here.

You can also email us. Opening your email app does not send an enquiry automatically.

WeChat · Nastia

English-speaking contact

WeChat QR code: Nastia

Scan the QR code to add Nastia on WeChat.